Bacharach Flue-Gas Sample Conditioning System Especificaciones Pagina 54

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 86
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 53
Calibration ECA 450
50 Instruction 0024-9400
3.2 Calibration Fixtures
A gas and pressure fixture will be required to perform the various calibration
procedures described in the sections that follow.
Parts Required:
Calibration Kit (refer to Section 6.2)
Calibration Gas (refer to Section 6.2)
Bellows
Micro manometer
Procedure:
Assemble the appropriate fixture as required by the calibration procedure
being performed. See Figure 3-1.
Figure 3-1. Calibration Fixtures
OPT T-STACK T-AIR
PRESSUREGAS
P REF
ECA 450
MENU
PRINT
I/O
RUN
SAVE
1
2 3 4 5
Leave Top
Port Open
3
3
GAS
FIXTURE
GAS
INLET
PORT
RS 232
POWER
OPT T-STACK T-AIR
PRESSUREGAS
P REF
ECA 450
MENU
PRINT
I/O
RUN
SAVE
PRESSURE
INLET
PORT
Parts Shown:
1. Gas Cylinder
2. Regulator*
3. Tubing*
4. Tee*
5. Flowmeter*
6. Fitting, Gas*
7. Fitting, Pressure*
* Contained in Calibration Kit
3
4
7
3
Micro manometer
Bellows
6
PRESSURE
FIXTURE
Vista de pagina 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 85 86

Comentarios a estos manuales