ECA 450 Calibration
Instruction 0024-9400
49
3 Calibration
Bacharach recommends that the ECA 450 be calibrated by your nearest
Bacharach Service Center (refer to Section 6.3). The analyzer, however, can
be calibrated in the field if your facility has the necessary equipment and
qualified personnel to perform the procedures described in this section.
Important! To prevent the loss of data during calibration, perform the
following procedures with a fully charged battery pack, or with the AC
adapter connected to the analyzer.
3.1 Initial Sensor Check
Important! Before turning on the analyzer or performing any of the calibration
procedures, ensure that the analyzer will be sampling fresh air and the probe
is at room temperature.
When the analyzer is first turned on and allowed to cycle through its
60 second warm-up period, and while sampling fresh air, the sensors are
checked (read) and calibrated (set) to the following ambient conditions:
Oxygen sensor spanned to 20.9%
All remaining gas sensors are zeroed
Pressure sensor zeroed
If a sensor is found to be in error during warm up, an error message will
appear at the bottom of the screen describing the nature of the problem (refer
to Section 5.4 Troubleshooting Guide). Correct any sensor errors before
proceeding with these calibration procedures.