Operation Fyrite
INSIGHT
Plus Manual
64 0024-9487 Rev 2
4.3. Combustion Testing Process
WARNING: The Fyrite
INSIGHT
Plus calculates combustion
parameters based on North American or Siegert combustion
equations. NA or Siegert configuration is selected in the SETUP
MENU. Be sure that your Fyrite
INSIGHT
Plus is properly
configured for your region and desired combustion calculations.
Step Example Combustion Testing Procedure
1 Confirm that testing prerequisites have been completed (see page 61).
2 Based on the sampling point examples (see page 61) and your combustion
application, locate and prepare an appropriate sampling point.
3 Insert the probe into the combustion location.
4 Press the RUN/HOLD button to begin sampling gas. You should see the
word RUN in the upper left corner of the display and hear the sample
pump turn on. If you see the word HOLD, press the RUN/HOLD button
again.
5 Monitor the display for combustion data.
6 If desired, turn on your optional IrDA printer, then press the F1 button on
the Fyrite
INSIGHT
Plus to print the current combustion data. (See
page 69 for additional printing information.)
7 Press the F3 button as desired to save combustion data for later retrieval,
review, and/or printing.
8 Press the RUN/HOLD button to stop the test. You should see the word
HOLD in the upper left corner of the display and hear the sample pump
turn OFF. If you see the word RUN, press the RUN/HOLD button again.
(You may optionally choose to print test data while in HOLD mode.)
9 Remove the probe from the sampling point and disconnect the probe.
CAUTION: The probe may be very hot. Allow it to cool,
then wipe it clean with a dry cloth.
10 Move the instrument to a clean air environment and press the POWER
button to turn off the instrument. The shutdown procedure includes a
purge component that clears the sensors of combustion gases.