Plus Manual Configuration
0024-9487 Rev 2 37
Setup Menu Function
The measured value of CO can be referenced to a
specific O
2
percentage (0% to 15%) as referenced in the
equation below.
𝐶𝑂(𝑛) =
20.9 𝑂
2
𝑅𝑒𝑓𝑒𝑟𝑒𝑛𝑐𝑒
20.9 𝑂
2
𝑀𝑒𝑎𝑠𝑢𝑟𝑒𝑑
× 𝐶𝑂
Use the UP (
) and DOWN (
) arrow buttons to enter
the O
2
reference value (n) from 0% to 15%. Press
ENTER to save the selection or ESC to revert to the
previous setting.
NOTE: The O
2
reference has a default
value of 0%. CO with respect to a 0% O
2
reference is also known as CO Air Free or
CO(0).
Select whether to print (YES) or not print (NO) the
pressure measurement
on the combustion test
printout.
Use the UP (
) and DOWN (
) arrow buttons to select
YES or NO. Press ENTER to save your choice.