Plus Manual Configuration
0024-9487 Rev 2 55
3.12. Calibration Menu
Calibration Menu Function
Calibration is performed by applying known values and
accessing the password-protected menu items. When
the Calibration Menu is selected, the user must enter a
4-digit numeric security code in order to proceed to
the calibration options. The default password is 1111.
Use the UP () and DOWN () arrow buttons to scroll
through numerals 0-9 until the desired numeral is
reached. Press ENTER to advance to the next position
of the password. Press ENTER after all four digits are
set. Press ESC to return to the SETUP MENU.
NOTE: The calibration password can be
changed through the Fyrite
User
Software (FUS).
Refer to Chapter 5 for additional screens and
calibration procedures.
• Pressure Calibration ............................ see page 86
• T-Stack Calibration .............................. see page 88
• T-Air Calibration .................................. see page 90
• CO Calibration ..................................... see page 93
• T-Ref Calibration .................................. see page 95
• B-SMART
Calibration ......................... see page 85