Bacharach PCA3 Especificaciones Pagina 80

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 79
PCA
3 Manual
80 0024-9473 Rev 4
6.11. CO
HIGH
Sensor Calibration
This procedure spans the optional carbon monoxide high sensor (4,001 to
20,000 ppm) to a known gas level.
6.11.1. Material Required
Calibration Kit (P/N 0024-7059)
Gas Cylinder (4,000 ppm CO in Air, P/N 0024-1155)
Figure 6-12. Gas Sensor Calibration Equipment
6.11.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
2. Use the ▲▼ buttons to highlight CO-HI, and then press ENT to display
the CALIBRATE CO-HI screen.
NOTE:
“Measured” is the current CO reading, while
“Applied” is a known CO level that will be applied for
calibration purposes.
3. Attach a 4,000 ppm CO cylinder to the regulator of the calibration fixture
shown in Figure 6-12.
4. Use the ▲▼ and
◄►
buttons to enter an applied value that exactly
equals the concentration on the CO cylinder.
Vista de pagina 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 103 104

Comentarios a estos manuales

Sin comentarios