Bacharach PCA3 Especificaciones Pagina 77

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 76
PCA
3 Manual
0024-9473 Rev 4 77
NOTE: If the sensor’s output is low, but still usable, then the
message “Good Calibration
WARNING Low Sensor” will
appear. The sensor will now be marked as being Low in the
DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then the
message “Bad Calibration Sensor End of Life, Entry Not
Saved” will appear followed
by the CALIBRATION LIST
screen being re-displayed. The sensor will now be marked as
being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.
6.9. NO Sensor Calibration
This procedure spans the optional nitric oxide sensor to a known gas
level.
6.9.1. Material Required
Calibration Kit (P/N 0024-7059)
Gas Cylinder (250 ppm NO in Nitrogen, P/N 0024-1156)
6.9.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
2. Use the ▲▼ buttons to highlight NO, and then press ENT to display the
CALIBRATE NO screen.
NOTE: “Measured” is the current NO reading, while “Applied”
is a known NO level that will be applied for calibration
purposes.
3. Attach a 250 ppm NO cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 80.
4. Use the ▲▼ and
◄►
buttons to enter an applied value that exactly
equals the concentration stamped on the NO cylinder.
NOTE: The calibration range is from 9 to 900ppm. An attempt
to calibrate outside this range will cause the message “Bad
Calibration Wrong CAL Entry” to appear in the following step.
Vista de pagina 76
1 2 ... 72 73 74 75 76 77 78 79 80 81 82 ... 103 104

Comentarios a estos manuales

Sin comentarios