Bacharach PCA3 Especificaciones Pagina 54

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 53
PCA
3 Manual
54 0024-9473 Rev 4
Figure 5-15. Emptying the Water Trap
5.12. Data Logging
5.12.1. Overview
When the logging function is activated, up to 500 combustion test records will be
automatically stored in memory at a preset interval over a predetermined
duration.
The maximum duration that data can be collected is determined by the interval.
For example, if the interval is set to 10 seconds, then the maximum selectable
length of time in which data can be collected to fill 500 memory locations would
be 1
hour (500 x 10 seconds = 5000 seconds or 83
minutes). If the operator
chooses a duration that is longer than possible for the interval chosen, then the
analyzer automatically selects the highest duration possible for the selected
interval.
The stored logged data can either be viewed on the display using the analyzer’s
memory function (see Memory Functions on page 58), or downloaded to a
personal computer using the supplied Fyrite
®
User Software and USB cable.
Downloaded data is stored on the computer’s hard drive as a comma separated
value ASCII text file with a CSV extension, which can be opened by most
spreadsheet programs for analysis.
Vista de pagina 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 103 104

Comentarios a estos manuales

Sin comentarios