PCA
3 Manual
0024-9473 Rev 4 59
Figure 5-19. Recalling Combustion Test Data
5.13.3. Recalling Logged Test Data
Recall individual logged combustion test data records as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight MEMORY, and then press ENT to
display the MEMORY MENU.
3. Use the ▲▼ buttons to highlight LOGGING DIRECTORY, and then
press ENT to display the LOGGING DIRECTORY screen.
4. Use the ▲▼ buttons to highlight the desired data logging location to be
recalled. Each logging location is identified by the date and time at
which the data was saved. The word EMPTY signifies that the logging
location does not contain data.
5. Press ENT to display the logging data contained in the selected memory
location.
NOTE:
The top line of the log test data screen shows the log
location being viewed. In this example, “Log: 1” is being displayed.
NOTE: The recalled logged test data can be printed by pressing the
PRINT (F1) button. Refer to Printing Test Data section on page 63.
Figure 5-20. Recalling Logged Test Data