PCA
3 Manual
0024-9473 Rev 4 67
5. Before calibration can begin a 4-character alphanumeric password must
be entered. Use the ▲▼ and ◄► buttons to enter the password. (The
default password is 1111.)
6. Press the ENT key to accept the password. If the correct password was
entered the CALIBRATION LIST Menu will be displayed.
7. Use the ▲▼ buttons to highlight the sensor to be calibrated, and then
perform the calibration procedure for that sensor as described in the
following sections.
Figure 6-1. Starting a Calibration
6.3. B-Smart
®
Sensor Replacement and Calibration
To replace a B-Smart
®
sensor and calibrate, do the following.
1. Enter the CALIBRATION MENU.
2. Use the ▲▼ buttons to select B-SMART
®
. Press ENTER to display the
B-Smart
®
code screen.
3. Use the ▲▼ buttons to enter the 10- or 14-digit alphanumeric code
supplied with the B-Smart
®
sensor. Use the ◄► buttons to move the
cursor across the screen. Press ENTER.
NOTE: There are two B-Smart
®
sensor code lengths: 10 digits for
CO
HIGH
, NO, NO
2
, and SO
2
, and 14 digits for CO
LOW
.
NOTE: If the correct code was entered, the analyzer accepts it and
returns to the CALIBRATION MENU. If an incorrect code was
entered, the screen will display “Invalid Code”. Check to make sure
the correct code has been entered. If the problem persist, contact
your nearest Bacharach Service Center.
NOTE: To save time, B-Smart
®
codes can be entered using the
Fyrite User Software.