Appendix PEM-9004
Teledyne Analytical Instruments 60
Accuracy: ± 5 ppm (up to 150 ppm)
± 5% of reading (up to 2000 ppm)
Response Time (T90): <60 sec
CO Measurement _____________________________
(without H2 compensation)
(Option)
Sensor: Electrochemical, electrochemical cell
Range: 0-2.0% vol (20,000 ppm)
Resolution: 0.01% vol
Response Time (T90): <60 sec
SO2 Measurement _____________________________
(Option)
Sensor: Electrochemical, electrochemical cell
Range: 0-2000 ppm
Resolution: 1 ppm
Accuracy: ± 5 ppm (up to 150 ppm)
± 5% of reading (up to 2000 ppm)
Response Time (T90): <60 sec
Waste or Flue Gas Temperature Measurement _____________
Sensor: K-type thermocouple (NiCr-Ni)
Range: + 32—1850°F (0—1000°C)
Resolution: 0.1°F or °C
Accuracy: ± 2°F, ± 1°C (0—400°C)
±0.5% of reading (up to 1000°C)
Combustion Air or Ambient Air
Temperature Measurement _____________________________
Sensor: K-type thermocouple (NiCr-Ni)
Range: -5—212°F (-20—100°C)
Resolution: 0.1°F or °C
Accuracy: ± 2°F, ± 1°C (0—100°C)
± 6°F, ± 3°C (-20—0°C)