Product Leadership • Training • Service • ReliabilityInstruction 2100-9001Operation and MaintenanceRev. 0 — June 2010

Pagina 2 - WARRANTY

Pur•Chek™ ProInstruction 2100-900163.3 High Side Liquid Sample HoseThe 6 foot (1.8 meter) High Side Sample Hose is constructed of a polyurethane tube

Pagina 3 - Contents

Pur•Chek™ ProInstruction 2100-9001 73.6 Back Panel ConnectionsThe connections located on the back panel are illustrated below.CAUTION: The sample out

Pagina 4 - Instruction 2100-9001ii

Pur•Chek™ ProInstruction 2100-900184 Operation4.1 Battery InstallationThe Pur•Chek™ Pro has, as an option, an internal rechargeable battery. If your

Pagina 5 - 1 Introduction

Pur•Chek™ ProInstruction 2100-9001 94.2 Turning On the UnitConnect the included AC power supply to the 12VDC power input jack on the back of the unit

Pagina 6 - Instruction 2100-90012

Pur•Chek™ ProInstruction 2100-9001104.4 Vapor SamplingVapor sampling is the most common process used for identifying refrigerants using the Pur•Chek™

Pagina 7 - 1.5 Functional Overview

Pur•Chek™ ProInstruction 2100-9001 11If the refrigerant tested is 90% pure or better, the results will be displayed as shown in Figure 9. Should the r

Pagina 8 - 2 Specifi cations

Pur•Chek™ ProInstruction 2100-9001124.9 Printing the Test ResultsThe test results can be printed by selecting the “PRINT” button. After the print is

Pagina 9 - 3 Components

Pur•Chek™ ProInstruction 2100-9001 135 Maintenance & Troubleshooting5.1 Setting the ElevationDuring the initial power-up, the Pur•Chek™ Pro will

Pagina 10 - 3.5 Control Panel

Pur•Chek™ ProInstruction 2100-9001145.3 Changing the White Plastic Sample FilterWhen inspecting the sample fi lter, look completely around the entire

Pagina 11 - 3.6 Back Panel Connections

Pur•Chek™ ProInstruction 2100-9001 15 TO CHANGESETTINGS ORREPLACE PAPERROLL PRESS SET FOR OTHER HELPPRESS HELPFigure 22.Figure 23.SET HELP

Pagina 12 - 4 Operation

WARRANTYBacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mate-rial and manufacture and will co

Pagina 13 - 4.3 Calibration

Pur•Chek™ ProInstruction 2100-9001166 Replacement PartsPart Number Description2100-0003 Printer Paper (Single Roll)2100-0004 AC Adapter2100-0006 Repla

Pagina 14 - 4.5 Liquid Sampling

Pur•Chek™ ProInstruction 2100-9001 17NOTES:

Pagina 15 - 4.8 Pure Blend Refrigerants

World Headquarters:621 Hunt Valley Circle, New Kensington, PA 15068PH: 724-334-5000 • Toll Free: 800-736-4666 • Fax: 724-334-5001Website: www.mybachar

Pagina 16 - Instruction 2100-900112

Pur•Chek™ ProInstruction 2100-9001 iContents1 Introduction ...11

Pagina 17 - 5.2 Setting the LCD Contrast

Pur•Chek™ ProInstruction 2100-9001ii

Pagina 18 - Instruction 2100-900114

Pur•Chek™ ProInstruction 2100-9001 11 Introduction1.1 How to Use This ManualThis manual provides important information on how to operate and maintain

Pagina 19 - INCORRECT

Pur•Chek™ ProInstruction 2100-900121.2.2 General Warnings & Cautions • Always inspect the sample hose before each use. Replace the hose if it a

Pagina 20 - 7 Service Centers

Pur•Chek™ ProInstruction 2100-9001 31.5 Functional OverviewContamination of refrigerants either in storage cylinders or air conditioning systems can l

Pagina 21 - Instruction 2100-9001 17

Pur•Chek™ ProInstruction 2100-900142 Specifi cationsSample Parameters Vapor only, oil-free, 500psig max.Identifi ed Refrigerants R408A, R409A, R417A, R4

Pagina 22 - World Headquarters:

Pur•Chek™ ProInstruction 2100-9001 53 Components3.1 Base UnitThe Pur•Chek™ Pro base unit houses the Graphic Display, Infrared Bench, Electrical Connec