PCA
3 Manual
0024-9472 Rev 3 85
Figure 7-7. Sensor Assembly
12. Allow the sensor that was just installed time to stabilize in the circuit
before continuing with this procedure. Stabilization time for all sensors
(except for the NO sensor) is about 1 hour. The NO sensor baseline
technically requires several days to stabilize, but should be sufficiently
stabilized for use in approximately 4 hours.
13. Turn ON the analyzer and confirm that no sensor errors occur during
warm up.
NOTE: Discard old sensors in accordance with local and federal
hazardous waste disposal laws.
14. Enter the B-Smart
®
calibration code or calibrate the new sensor(s) as
explained earlier (see Calibration on page 62). An exception is the O
2
sensor which does not require calibration.
CAUTION: