Instruction 0011-9026 Page 1Printed in U.S.A.® Registered TrademarksINSTRUCTION 0011-9026FYRITE® Gas AnalyzerCO2 and O2 IndicatorsOperation/Maintenan
Instruction 0011-9026Page 67. Absorb sample gas into FYRITE by inverting until fl uid drains into top reservoir (Fig. 7). Then turn upright (Fig. 8)
Instruction 0011-9026 Page 710. This step completes CO2 or O2 FYRITE gas sample analysis. A few points to re-member when reading the FYRITE:(a) FYR
Instruction 0011-9026Page 83.2.1 Draining Fluid DropletsFor maximum accuracy, it is important to form the habit of follow ing a standard procedure
Instruction 0011-9026 Page 9Figure 12. When plunger valve is depressed, a gas sample is pumped through top reservoir with center bore is sealed off.
Instruction 0011-9026Page 10To make a test with the FYRITE, the metal Sampling Tube at one end of Sampling Assembly Hose is inserted into the gas to
Instruction 0011-9026 Page 11FYRITE CO2 refi ll fl uid may develop a white insoluble precipitate upon aging. The precipitate does not affect fl uid perf
Instruction 0011-9026Page 120-7.6% CO2 or O2CO2 tests of controlled atmospheres in fruit, vegetable, and meat storage rooms. Oxygen determination in
Instruction 0011-9026 Page 13Excess combustion air is heated and carries some of this heat to the fl ue where it is wasted. The FYRITE CO2 Analyzer is
Instruction 0011-9026Page 14Proper O2 content for any fuel fi red is the lowest O2 value that will ensure complete, clean combustion with adequate sa
Instruction 0011-9026 Page 15Figure 15. Relation between Oxygen, CO2 and excess air in fl ue gases for Natural Gas and Fuel Oil.THEORETICAL EXCESS AI
WARRANTYBacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will c
Instruction 0011-9026Page 16As a rule of thumb, background gases/vapors may be present in con centrations up to 1/2% by volume (5000 ppm) before the
Instruction 0011-9026 Page 17 Figure 17. Checking CO2 of gas designed furnace (sampling tube inserted through draft diverter;
Instruction 0011-9026Page 18The gas concentration read on the FYRITE is directly dependent upon the mass of air in the sample. The aspirator bulb us
Instruction 0011-9026 Page 19Refer to Figure 19. To remove excess fl uid, insert small diameter glass tube into FY-RITE fl uid through the small center
Instruction 0011-9026Page 20®20161284%CO2020161284%CO20®201612840®%CO2TOP GASKETTOP CAP/PLUNGERVALVE ASSEMBLYOVAL HEADSCREWSTOP CAP RINGWITH GASKET2
Instruction 0011-9026 Page 21accuracy don’t forget to saturate fi lter saturator wool packing as such gas standards are usually supplied “bone dry”.6.
Instruction 0011-9026Page 22Drain old fl uid from FYRITE and rinse all parts in clean, lukewarm water.NOTE: FYRITE fl uid is corrosive to skin, clothi
Instruction 0011-9026 Page 2320161284%C O2020161284%CO20®20161284%CO20®GASKETBODYTOP CAPASSEMBLYTop Gasket used in 7%, 20% & 21% Models only®2016
Instruction 0011-9026Page 24mine replacement parts necessary for repair, or drain instrument and return if factory service or repair is desired.6.5
Instruction 0011-9026 Page 25Figure 25a. Replacement Diaphragm properly installed.Figure 25. Locating Bezel Rubber Gaskets.Figure 26. FYRITE Sampl
FYRITE SAFETY PRECAUTIONSThe operator (s) of this instrument should thoroughly familiarize themselves with the applicable safety precautions before ha
Instruction 0011-9026Page 26with fi nger and squeeze Aspirator Bulb (Figure 27). Bulb should remain fi rm. If Bulb collapses, check Bulb and Hose to C
Instruction 0011-9026 Page 27ASPIRATOR BULB SHOULDREMAIN FIRM WHEN SQUEEZEDRUBBER CONNECTOR TIP(HOLD FINGER HERE)CHECK FOR BULB INFLATIONAFTER SQUEEZ
Instruction 0011-9026Page 28Figure 30. Wrapping Replacement Filter Material.Figure 31. Wet Filter Material then squeeze out excess water.Figure 32
Instruction 0011-9026 Page 29Part # Description Qty-Rq’d 11-0102 Bezel 411-0188 Oval Head Screw 811-0
Instruction 0011-9026Page 307.1 FYRITE ILLUSTRATED PARTSFigure 33. FYRITE parts breakout.11-001911-002611-018811-018811-010211-013611-010911-0132
Instruction 0011-9026 Page 310011-01060011-01180011-01180011-0122(0011-0180 + 0011-0130)0011-01190011-0138 0011-01380011-01200011-01527.2 PARTS LI
Instruction 0011-9026Page 32Current FormerPart # List # Description # Req'd11-0152 10-0019 Connector Tip with Tube
Instruction 0011-9026 Page 33Notes
Instruction 0011-9026Page 34Notes
Instruction 0011-9026 Page 35Notes
TABLE OF CONTENTS1.0 DESCRIPTION...
Instruction 0011-9026Page 36Bacharach, Inc.621 Hunt Valley Circle, New Kensington, PA 15068Phone: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-
Instruction 0011-9026 Page 1INSTRUCTIONS 0011-9026OPERATION/MAINTENANCEBACHARACH FYRITE MODELSCO2 and O2 INDICATORS1.0 DESCRIPTION (Refer to Figure
Instruction 0011-9026Page 2(b) FYRITE Fluid Strength To check fl uid strength.NOTE: When repeating procedures as outlined in Section 3.0 Steps 7 thr
Instruction 0011-9026 Page 3HAZARDSCorrosive liquid causes burns. May cause blindness if splashed in eyes. Vapors are irritating and may be harmful.P
Instruction 0011-9026Page 4CO2 FYRITE OPERATION O2 FYRITE OPERATION (RED FLUID) (BLUE FLUID)1. Hol
Instruction 0011-9026 Page 5Figure 5Figure 5aFigure 5bFigure 6CO2 FYRITE OPERATION O2 FYRITE OPERATION (RED FLUID)
Comentarios a estos manuales